13 Unit adjustment and calibration
Sirona Dental Systems GmbH
13.2 Adjustment and calibration via the calibration menu
Installation Manual ORTHOPHOS SL
4. Repeat the procedure starting at Step 2 until the values in the menu's
text boxes are equal to "zero". This means that adjustment was
performed successfully.
IMPORTANT: If automatic adjustment fails to produce values equal
to zero, calculate the adjustment values manually and overwrite the
values in the menu's text boxes with these values [ → 225].
5. If adjustment is OK (the adjustment values in the text boxes are equal
to "zero"), save the values. [ → 209]
The status indicator that prefixes the
"Diaphragm adjustment"
element is checked.
6. Continue with the next stage of the adjustment procedure [ → 227].
A
Adjustment OK
● The exposed diaphragm area must lie
horizontally centered in the image
section as well as inside the auxiliary
lines.
● A surrounding white border must be
present.
● The maximum density must lie in the
center of the diaphragm area.
B
Adjustment not OK