13 Unit adjustment and calibration
Sirona Dental Systems GmbH
13.2 Adjustment and calibration via the calibration menu
Installation Manual ORTHOPHOS SL
13.2.1.6
Adjustment of the ceph secondary diaphragm
13.2.1.6.1
Automatic adjustment: Ceph secondary diaphragm (standard)
Phantom for Ceph exposure
✔
The pan needle phantom is
not
in the pan bite block holder on the
unit.
✔
The ceph test phantom does
not
connect to the cover of the sensor
unit.
IMPORTANT
If you cannot correctly adjust the ceph secondary diaphragm via the
calibration menu, the Ceph secondary diagram must first be readjusted
After mechanical readjustment has been completed, repeat adjustment
via the calibration menu.
NOTICE
Risk of damage to unit
It is essential that the needle phantom is removed from the bite block
holder on the unit before a ceph exposure is taken; otherwise, the
sensor could collide with the phantom.