background image

INSTRUCTION MANUAL

Read and understand all of the instructions and safety

information in this manual before operating this product.

MAN-028 Rev B

08/2021

© 2021 MAC Medical, Inc.

Add-on Scrub Sink Timer Kit

for Installation on Economy Processing Sinks

Summary of Contents for S0202-EST-KT

Page 1: ...d and understand all of the instructions and safety information in this manual before operating this product MAN 028 Rev B 08 2021 2021 MAC Medical Inc Add on Scrub Sink Timer Kit for Installation on Economy Processing Sinks ...

Page 2: ...llation and Timer setup programming instructions for the devices contained in this kit Table of Contents Parts List for S0202 EST KT 3 Plumbing Installation 4 Timer S0050 Specifications and Terminal Connections 7 Terminal Wiring Connections and Battery Replacement 8 Timer Operation Display and Programming 9 InLine Flow Switch Timer Controller S0130 01 Specifications Switch Rating and Wiring Diagra...

Page 3: ...tion QTY 1 S0050 Timer Eagle 1 2 S0130 01 Inline Flow Switch Timer Control Brass 1 3 S0090 01 Timer Housing Weldment 1 4 S0019 Union 1 2 Copper 3 5 S0049 Elbow 1 2 Copper 90 4 6 S0165 Pipe Fitting Reducer 1 2 Sweat X 1 4 Npt Male 2 7 S0069 1 2 Copper Hard Type L Cierro Brand 10 Links Only 3 ft ...

Page 4: ... inches below elbow see red cut line New Unions New Union Flow Switch Timer After the cuts the original pipe will be relocated on the sink The sketch at left shows the new plumbing configuration outlined in black and showing unions and timer location In the new plumbing configuration use the kit plumbing items flow switch timer two reducers three unions four elbows additional pipe length to rerout...

Page 5: ...iginal plumbing straightness S0049 Elbows S0165 Reducers top and side S0130 01 Timer S0019 Unions S0049 Elbow S0019 Union S0049 Elbow The cut original pipe is now located to the left from its previous position 18 1 2 length pipe S0069 or trim to allow for elbow 2 1 4 Length S0069 pipe 1 1 2 length pipes S0069 5 length pipe S0069 Detail See Detail inset below left ...

Page 6: ...in the illustration below Two holes are for mounting the timer housing one for threading the three flow switch timer wires to the top side for connection to the timer read out Place a grommet in the wire threading hole Grommet to be supplied by others Mount the timer housing to the top of the sink in the 0 31 diameter drilled holes Mounting studs are provided on the housing Secure to sink top with...

Page 7: ...w single battery 10 years w dual batteries Ranges Resolution Seconds minutes to 1 10 hours to 1 10 hours minutes seconds Display 12 mm high Supertwist LCD 7 digits Low Bat indicator Backlighting Green Illumination over whole viewable area Requires 10 to 28 VDC power source Dimensions Mounting See dimensions picture below Panel Mount with supplied mounting bracket and gasket Connections Screw termi...

Page 8: ...be run on two batteries to extend the battery life to 10 years Once the battery is in place the unit will go into a self test mode and all the segments on the LCD display will be illuminated The self test mode is exited by depressing the Next key which will then display the model number 6 Depress the Next key again to ready the unit for operation Terminal Wiring Connections and Battery Replacement...

Page 9: ...ing parameters can be accessed when the Program Enable input is active by pressing the Down key To edit a parameter use the Down key to scroll until the desired parameter appears on the screen Pressing the Next key will cause the leftmost digit of that value to begin to flash Use the Next and Down keys in combination to choose individual digits and change their value Front Panel Reset Enable When ...

Page 10: ...h Rating and Wiring Diagram This is an explosion proof brass flow switch actuation set point 0 50 GPM 1 89 LPM and calibrated for water at standard conditions It is used for accurate detection of excessive or insufficient flow rates Switch Ratings Max Resistive Load This unit was calibrated in a vertical position with lead wires up Install unit in piping system using standard pipe fitting procedur...

Page 11: ...ally be removed from these switches Occasional wipe down cleaning when excessive contamination is present is all that is normally required To Clean Remove unit from system and disassemble as shown below Clean all parts reassemble and reinstall unit Note 50 micron filtration is recommended Piston Inlet Fitting Housing Spring O Ring ...

Page 12: ...ll Free 877 828 9975 www macmedical com MAC Medical Inc Printed in USA Publication No MAN 028 Rev B August 2021 Information regarding this product is subject to change without prior notice Customer needs are our first priority To place an order contact our customer service department at 618 476 3550 or 877 828 9975 or by email at sales macmedical com ...