[HEATER VOLTAGE SETTING ERROR]
This error appears when the voltage switches are set to 115V and
the voltage of 230V is applied to the Heaters because the heater
temperature experienced a sudden rise.
Voltage : 230V
Voltage Switches : 115V
[HEATING TIMEOUT CONTINUE?]
This error appears when the heaters didn’t reach the set
temperatures even if approximately 20 minutes had passed.
It takes a long time to reach the set temperature and this error
may appear when the voltage switches are set to 230V
(Default)
and the voltage of 115V is applied to the Heaters.
Voltage : 115V
Voltage Switches : 230V
Features_Heater
Features_Heater
Error
- Long time to reach
the temp.
- Error