Remote Control
R&S
®
FSPN
391
User Manual 1179.4363.02 ─ 01
Manual operation:
See
"Defining the video bandwidth"
11.6.9.4
Trigger Configuration
...........................................................................................391
...................................................................................391
TRIGger[:SEQuence]:FREQuency:HYSTeresis
................................................................. 391
TRIGger[:SEQuence]:HOLDoff[:TIME]
..............................................................................392
TRIGger[:SEQuence]:IFPower:HOLDoff
........................................................................... 392
TRIGger[:SEQuence]:LEVel[:EXTernal<1|2>]
.................................................................... 392
.......................................................................................... 393
........................................................................................393
TRIGger[:SEQuence]:DTIMe
<Time>
This command defines the trigger drop-out time.
Parameters:
<Time>
<numeric value>
*RST:
0
Default unit: s
Example:
//Define a drop-out time
TRIG:DTIM 200MS
Manual operation:
See
"Defining a trigger drop-out time"
TRIGger[:SEQuence]:FREQuency
<Frequency>
This command defines the trigger frequency.
Prerequisites for this command
●
Select the frequency trigger (
).
Parameters:
<Frequency>
<numeric value>
Default unit: Hz
Example:
//Configure frequency trigger
TRIG:SOUR FREQ
TRIG:FREQ 1GHZ
Manual operation:
See
"Defining the trigger level and frequency"
TRIGger[:SEQuence]:FREQuency:HYSTeresis
<Hysteresis>
This command defines the trigger hysteresis.
Prerequisites for this command
●
Select the frequency trigger (
).
Common Measurement Settings