Remote Control
R&S
®
FSPN
480
User Manual 1179.4363.02 ─ 01
Parameters:
<Bandwidth>
0.1 Hz | 1 Hz | 3 Hz | 10 Hz | 30 Hz | 100 Hz | 300 Hz
The possible values depend on the reference source and tuning
range (see
Default unit: Hz
Example:
ROSC:LBW 3
Manual operation:
See
[SENSe:]ROSCillator:O640
<State>
This command turns the output of a reference signal on the corresponding connector
("Ref Output") on and off.
[SENSe:]ROSCillator:O640
: Provides a 640
MHz reference signal on correspond-
ing connector.
Parameters:
<State>
ON | OFF | 1 | 0
OFF | 0
Switches the reference off.
ON | 1
Switches the reference on
Example:
//Output reference signal of 100 MHz.
ROSC:O100 ON
Manual operation:
See
SOURce<si>:EXTernal<ext>:ROSCillator:EXTernal:FREQuency
<Frequency>
This command defines the frequency of the external reference oscillator.
If the external reference oscillator is selected, the reference signal must be connected
to the rear panel of the instrument.
Suffix:
<si>
.
1..n
<ext>
1..n
Example:
ROSC:EXT:FREQ 13MHZ
Sets the frequency to 13 MHz.
SOUR:EXT:ROSC:EXT:FREQ 13MHZ
Manual operation:
See
[SENSe:]ROSCillator:OSYNc
<State>
If enabled, a 100 MHz reference signal is provided to the "SYNC TRIGGER OUTPUT"
connector.
Configuring the R&S