Remote Control
R&S
®
FSPN
364
User Manual 1179.4363.02 ─ 01
[SENSe:]LIST:BWIDth[:RESolution]:USMall
<State>
Provides functionality to enable ultra-small resolution bandwidths (RBW) for the current
measurement.
Parameters:
<State>
ON | OFF | 1 | 0
*RST:
OFF
Example:
SENS:LIST:BWID:RES:USM ON
Manual operation:
See
[SENSe:]SWEep:COUNt
<SweepCount>
This command defines the number of measurements that the application uses to aver-
age traces.
In continuous measurement mode, the application calculates the moving average over
the average count.
In single measurement mode, the application stops the measurement and calculates
the average after the average count has been reached.
Parameters:
<SweepCount>
When you set a sweep count of 0 or 1, the R&S
FSPN performs
one single measurement in single measurement mode.
In continuous measurement mode, if the sweep count is set to 0,
a moving average over 10 measurements is performed.
Range:
0 to 200000
*RST:
0
Example:
SWE:COUN 64
Sets the number of measurements to 64.
INIT:CONT OFF
Switches to single measurement mode.
INIT;*WAI
Starts a measurement and waits for its end.
Manual operation:
See
[SENSe:]SWEep:FORWard
<SweepForward>
This command selects the sweep direction.
Parameters:
<SweepForward>
ON
Measurement in forward direction.
OFF
Measurement in reverse direction.
*RST:
OFF
Common Measurement Settings