General Instrument Setup
R&S
®
FSPN
234
User Manual 1179.4363.02 ─ 01
During instrument start, the installed hardware is checked against the current firmware
version to ensure the hardware is supported. If not, an error message is displayed
( "Wrong Firmware Version" ) and you are asked to update the firmware. Until the firm-
ware version is updated, self-alignment fails. To see which components are not suppor-
ted, see the
.
The firmware on your R&S
FSPN may also need to be updated in order to enable addi-
tional new features or if reasons for improvement come up. Ask your sales representa-
tive or check the Rohde&Schwarz website for availability of firmware updates. A firm-
ware update package includes at least a setup file and release notes.
Before updating the firmware on your instrument, read the release notes delivered with
the firmware version.
Enter the name or browse for the firmware installation file and press the "Install" but-
ton.
Remote command:
How to Update the Instrument Firmware
1. Download the update package from the Rohde&Schwarz website and store it on a
memory stick, on the instrument, or on a server network drive that can be accessed
by the instrument.
2.
NOTICE!
Stop measurement. The firmware update must not be performed during a
running measurement.