Remote Control
R&S
®
FSPN
506
User Manual 1179.4363.02 ─ 01
REMote
Queries the error messages that occurred during remote opera-
tion.
Note: The remote error list is automatically cleared when the
R&S
FSPN is shut down.
Return values:
<SystemMessages> String containing all messages in the "System Messages" table.
<RemoteErrors>
<Error_no> | <Description> | <Command> | <Date> | <Time>
Comma-separated list of errors from the "Remote Errors" table,
where:
<Error_no>
: device-specific error code
<Description>
: brief description of the error
<Command>
: remote command causing the error
<Date>|<Time>
: date and time the error occurred
Usage:
Query only
SYSTem:ERRor[:NEXT]?
This command queries the most recent error queue entry and deletes it.
Positive error numbers indicate device-specific errors, negative error numbers are error
messages defined by SCPI. If the error queue is empty, the error number 0, "No error",
is returned.
For details on error queues see
Chapter 10.1.7, "Status Reporting System"
Usage:
Query only
SYSTem:FIRMware:UPDate
<Directory>
This command starts a firmware update using the *.msi files in the selected directory.
The default path is
D:\FW_UPDATE
command. To store the update files the
Only user accounts with administrator rights can perform a firmware update.
Setting parameters:
<Directory>
Example:
SYST:FIRM:UPD 'D:\FW_UPDATE'
Starts the firmware update from directory "
D:\FW_UPDATE
".
SYSTem:FORMat:IDENt
<IDNFormat>
This command selects the response format to the
Parameters:
<IDNFormat>
LEGacy
Format is compatible to R&S FSP/FSU/FSQ/FSG family.
Configuring the R&S
FSPN