Remote Control
R&S
®
FSPN
390
User Manual 1179.4363.02 ─ 01
11.6.9.3
Data Acquisition
Commands to configure data acquisition described elsewhere.
●
●
Sweep mode:
●
Sweep control:
●
Sweep control:
........................................................................................ 390
........................................................................................ 390
[SENSe:]AVERage<n>:COUNt
<AverageCount>
This command defines the number of measurements that the application uses to aver-
age traces.
In case of continuous sweep mode, the application calculates the moving average over
the average count.
In case of single sweep mode, the application stops the measurement and calculates
the average after the average count has been reached.
Suffix:
<n>
.
irrelevant
[SENSe:]BWIDth:VIDeo
<Bandwidth>
This command defines the video bandwidth.
Prerequisites for this command
●
Turn off automatic VBW selection (
Parameters:
<Bandwidth>
Default unit: HZ
Example:
//Select VBW manually
BWID:VID:AUTO OFF
BWID:VID 819672
Manual operation:
See
"Defining the video bandwidth"
[SENSe:]BWIDth:VIDeo:AUTO
<State>
This command turns automatic selection of the video bandwidth on and off.
Parameters:
<State>
ON | OFF | 1 | 0
*RST:
ON
Example:
//Turn on automatic VBW selection
BWID:VID:AUTO ON
Common Measurement Settings