Example:
//Select a frequency resolution of 100 Hz.
CONF:VCO:SWE:FCO 100HZ
Manual operation:
See
"Selecting the frequency resolution"
CONFigure:VCO:SWEep:POINts
<MeasPoints>
This command defines the number of measurement points for the VCO measurement.
Parameters:
<MeasPoints>
<numeric value> (integer only)
*RST:
10
Example:
See
Manual operation:
See
CONFigure:VCO:SWEep:SOURce
<PowerSource>
This command selects the source (connector) which tunes the VCO.
Parameters:
<PowerSource>
VAUX
Selects the V
aux
connector as the sweep source.
VSUPly
Selects the V
supply
connector as the sweep source.
VTUNe
Selects the V
tune
connector as the sweep source.
Example:
//Select the V
supply
connector as the tuning source
CONF:VCO:SWE:SOUR VSUP
//Select the type of output at the V
supply
connector
SOUR:VOLT:POW:LEV:MODE VOLT
//Configure a tuning range with 10 measurement points
CONF:VCO:SWE:STAR 0V
CONF:VCO:SWE:STOP 5V
CONF:VCO:SWE:POIN 10
//Turn on the output of the tuning signal.
SOUR:VOLT:POW:LEV ON
CONFigure:VCO:SWEep:STARt
<Value>
This command defines the start value of the tuning range.
Parameters:
<Value>
<numeric value>
The unit depends on whether you tune the VCO with a voltage
or current (see
*RST:
0
Default unit: V or A