[SENSe:]ADJust:CONFigure:LEVel:THReshold
<Level>
This command defines the threshold of the signal search.
Prerequisites for this command
●
Turn on the automatic signal search with
Parameters:
<Level>
<numeric value>
Default unit: dBm
Example:
//Define a threshold of -10 dBm
ADJ:CONF:LEV:THR -10
Manual operation:
See
"Searching for the signal level"
[SENSe:]POWer:RLEVel?
This command queries the currently measured signal level.
Return values:
<Level>
<numeric value>
Default unit: dBm
Example:
//Query signal level
POW:RLEV?
would return, e.g.
-9.28192
Usage:
Query only
Manual operation:
See
"Searching for the signal level"
[SENSe:]POWer:RLEVel:OFFSet
<Offset>
This command defines an arithmetic level offset. The offset is applied to the signal
level.
The level offset takes external attenuation or gain into account.
Parameters:
<Offset>
<numeric value>
Range:
-100 to 200
*RST:
0
Default unit: dB
Example:
//Define a level offset of 20 dB
POW:RLEV:OFFS 20
Manual operation:
See