Remote Control
R&S
®
FSPN
369
User Manual 1179.4363.02 ─ 01
Parameters:
<Frequency>
<numeric value>
Range:
Depends on the measurement range.
*RST:
1 kHz
Default unit: Hz
Example:
See
CALCulate<n>[:RANGe<j>]:EVALuation[:STATe]
Manual operation:
See
"Defining the integration range"
CALCulate<n>[:RANGe<j>]:EVALuation:TRACe
<Trace1>
This command selects the trace for calculation of residual effects.
Suffix:
<n>
.
<j>
Parameters:
<Trace1>
TRACE1 | ... | TRACE6
Selects the trace. You can select a trace that is inactive, but for
residual effect calculation, it must be active.
NONE
Turns off residual calculation for the respective range.
*RST:
Depends on the range.
Example:
//Use trace 1 as the data source for range 2
CALC:RANG2:EVAL:TRAC TRACE1
Manual operation:
See
"Selecting the trace for residual calculations"
CALCulate<n>[:RANGe<j>]:EVALuation[:STATe]
<State>
This command turns integration of the entire measurement range for residual calcula-
tions on and off.
Suffix:
<n>
.
<j>
Parameters:
<State>
ON | 1
Integrates over the complete measurement range.
OFF | 0
Integrates over a custom measurement range.
Example:
//Define an integration range from 1 kHz to 3 kHz in range 2
CALC:RANG2:EVAL OFF
CALC:RANG2:EVAL:STAR 100000
CALC:RANG2:EVAL:STOP 300000
Manual operation:
See
"Defining the integration range"
Common Measurement Settings