General Instrument Setup
R&S
®
FSPN
215
User Manual 1179.4363.02 ─ 01
Remote command:
9.1.3
How to Perform a Self-Test
You do not have to repeat the self-test every time you switch on the instrument. It is
only necessary when instrument malfunction is suspected.
Operating temperature
Before performing this functional test, make sure that the instrument has reached its
operating temperature (for details, refer to the data sheet).
1. Select [SETUP].
2. Select "Service" .
3. Select "Selftest" .
Once the instrument modules have been checked successfully, a message is dis-
played.
9.1.4
How to Align the Instrument
Operating temperature
Before performing this functional test, make sure that the instrument has reached its
operating temperature (for details, refer to the data sheet).
To perform a self-alignment
Make sure no signal is connected to the RF input connector. Running a self-alignment
with a signal connected to the RF input can lead to false measurement results.
1. Select [SETUP].
2. Select "Alignment" .
3. Select "Start Self Alignment" .
4. To abort the self-alignment process, select "Abort Self Alignment" .
Once the system correction values have been calculated successfully, a message
is displayed.
To display the alignment results again later
► Select [SETUP] > "Alignment" .
Alignment