Common Measurement Settings
R&S
®
FSPN
99
User Manual 1179.4363.02 ─ 01
Select an active window from the "Specific Settings for" selection list that is displayed
in the "Overview" and in all window-specific configuration dialog boxes.
The "Overview" and dialog boxes are updated to indicate the settings for the selected
window.
6.2
Input Source
The phase noise application supports input from several signal sources.
For more information about the available connectors, refer to the getting started man-
ual.
6.2.1
RF Input
Access
(RF input settings): "Overview" > "Input" > "Input Source" > "Radio Frequency"
> "Config"
Access
(schematic test setups): "Overview" > "Input" > "Input Source" > "Radio Fre-
quency" > "Test Setup"
The RF Input is the default input source.
A typical test setup for measurements over the RF input depends on the selected mea-
surement and the equipment used in the test setup. A schematic representation of
such a setup is provided in the dialog box.
●
The DUT directly sends a signal to the RF input of the R&S
FSPN.
●
If necessary, the R&S
FSPN can supply the DUT with power over the DC ports.
For more information about configuring the DC ports used in the setup, see
ter 6.6, "DC Source Configuration"
The remote commands required to configure the frequency are described in
................................................................................................ 99
Radio Frequency State
Activates input from the "RF Input" connector.
Input Source