Common Measurement Settings
R&S
®
FSPN
123
User Manual 1179.4363.02 ─ 01
The remote commands required to configure frequency stability measurements are
described in
Chapter 11.6.6.1, "Noise Configuration"
6.6
DC Source Configuration
Access
(all DC port settings): "Overview" > "Output" > "Output" > "DC Config"
Access
(most relevant DC port settings): (in the toolbar)
Access
(DC port state): "Overview" > (in the toolbar)
The "DC Power" settings allow you to configure the DC power supply sources available
on the front panel for different purposes.
Proper configuration of the DC ports
Setting DC voltages and currents properly is an important step during the configuration
in order not to damage the device under test (DUT) by applying too much voltage or
current!
Therefore it is recommended to connect the device under test (DUT) to the instrument
after the DC settings have been made and verified.
The remote commands required to configure the DC power output are described in
Turning the DC output on and off
................................................................................ 124
Coupling the output configuration across measurement channels
............................. 124
Configuring the DC power output
................................................................................124
........................................................................124
DC Source Configuration