Remote Control
R&S
®
FSPN
448
User Manual 1179.4363.02 ─ 01
Manual operation:
See
CALCulate<n>:LIMit<li>:UPPer:SPACing
<InterpolType>
This command selects linear or logarithmic interpolation for the calculation of an upper
limit line from one horizontal point to the next.
Suffix:
<n>
.
<li>
Parameters:
<InterpolType>
LINear | LOGarithmic
*RST:
LIN
Manual operation:
See
11.7.5.3
Reading out the Results of a Limit Check
CALCulate<n>:LIMit<li>:CLEar[:IMMediate]
...................................................................... 448
........................................................................................ 448
CALCulate<n>:LIMit<li>:CLEar[:IMMediate]
This command deletes the result of the current limit check.
The command works on
all
limit lines in
all
measurement windows at the same time.
Suffix:
<n>
.
<li>
irrelevant
Example:
CALC:LIM:CLE
Deletes the result of the limit check.
CALCulate<n>:LIMit<li>:FAIL?
This command queries the result of a limit check in the specified window.
To get a valid result, you have to perform a complete measurement with synchroniza-
tion to the end of the measurement before reading out the result. This is only possible
for single measurement mode.
Suffix:
<n>
.
<li>
Return values:
<Result>
0
PASS
Common Analysis and Display Functions