Network and Remote Operation
R&S
®
FSPN
309
User Manual 1179.4363.02 ─ 01
Restoring the connection to the R&S
FSPN
Follow the instructions above for setting up a connection to the R&S
FSPN. If the con-
nection is terminated and then restored, the R&S
FSPN remains in the same state.
10.4.7.4
How to Shut Down the R&S
FSPN via Remote Operation
1. Select the R&S
FSPN softfrontpanel and close the application with the key combi-
nation [ALT] + [F4].
2. Select the desktop and press the key combination [ALT] + [F4].
A safety query is displayed to warn you that the instrument cannot be reactivated
via remote operation and asks you whether you want to continue the shutdown
process.
3. Respond to the safety query with "Yes".
The connection with the controller is terminated and the R&S
FSPN is shut down.
10.4.8
How to Start a Remote Control Session from a PC
When you switch on the R&S
FSPN, it is always in manual operation state ("local"
state) and can be operated via the front panel.
To start remote control
1. Send an addressed command (
GTR
- Go to Remote) from a controller to the instru-
ment.
The instrument is switched to remote control ("remote" state). Operation via the
front panel is disabled. Only the "Local" softkey is displayed to return to manual
operation. The instrument remains in the remote state until it is reset to the manual
state via the instrument or via remote control interfaces. Switching from manual
operation to remote control and vice versa does not affect the other instrument set-
tings.
2. During program execution, send the
SYSTem:DISPlay:UPDate ON
command to
activate the display of results (see
The changes in the device settings and the recorded measurement values are dis-
played on the instrument screen.
3. To obtain optimum performance during remote control, send the
SYSTem:DISPlay:UPDate OFF
command to hide the display of results and dia-
grams again (default setting in remote control).
4. To prevent unintentional return to manual operation, disable the keys of the instru-
ment using the universal command
LLO
.
Switching to manual mode is only possible via remote control then. This function is
only available for the GPIB interface.