Remote Control
R&S
®
FSPN
326
User Manual 1179.4363.02 ─ 01
Example:
ABOR;:INIT:IMM
Aborts the current measurement and immediately starts a new
one.
Example:
ABOR;*WAI
INIT:IMM
Aborts the current measurement and starts a new one once
abortion has been completed.
Usage:
Event
INITiate<n>:CONMeas
This command restarts a (single) measurement that has been stopped (using
ABORt
)
or finished in single measurement mode.
The measurement is restarted at the beginning, not where the previous measurement
was stopped.
Suffix:
<n>
.
irrelevant
Example:
INIT:CONT OFF
Switches to single measurement mode.
DISP:WIND:TRAC:MODE AVER
Switches on trace averaging.
SWE:COUN 20
Setting the measurement counter to 20 measurements.
INIT;*WAI
Starts the measurement and waits for the end of the 20 mea-
surements.
INIT:CONM;*WAI
Continues the measurement (next 20 measurements) and waits
for the end.
Result: Averaging is performed over 40 measurements.
Manual operation:
See
INITiate<n>:CONTinuous
<State>
This command controls the measurement mode for an individual channel.
Note that in single measurement mode, you can synchronize to the end of the mea-
surement with *OPC, *OPC? or *WAI. In continuous measurement mode, synchroniza-
tion to the end of the measurement is not possible. Thus, it is not recommended that
you use continuous measurement mode in remote control, as results like trace data or
markers are only valid after a single measurement end synchronization.
Example:
INIT:CONT OFF
Switches the measurement mode to single measurement.
INIT:CONT ON
Switches the measurement mode to continuous measurement.
Measurements and Result Displays