Remote Control
R&S
®
FSPN
393
User Manual 1179.4363.02 ─ 01
Parameters:
<Level>
<numeric value>
Default unit: V
Example:
//Configure external trigger on trigger port 2
TRIG:SOUR EXT2
TRIG:LEV:EXT2 1.4V
Manual operation:
See
"Defining the trigger level and frequency"
TRIGger[:SEQuence]:SLOPe
<Slope>
This command selects the trigger slope.
Parameters:
<Slope>
NEGative
Triggering occurs on a falling signal edge.
POSitive
Triggering occurs on a rising signal edge.
*RST:
POSitive
Example:
//Select trigger slope
TRIG:SLOP POS
Manual operation:
See
TRIGger[:SEQuence]:SOURce
<Source>
This command selects the trigger source.
Parameters:
<Source>
EXTernal | EXT2
External trigger: The measurement starts when the signal
applied to the input connector meets or exceeds a certain level.
FREQuency
Frequency trigger: The measurement starts when the signal
applied to the RF input meets or exceeds a certain frequency.
IMMediate
Free run: No trigger is used.
IQPower
Power trigger: The measurement starts when the signal level
within the demodulation bandwidth rises or falls by a certain
amount.
*RST:
IMMediate
Example:
//Select trigger source
TRIG:SOUR FREQ
Manual operation:
See
"Selecting the trigger source"
Common Measurement Settings