Operating Procedures
PyroMark Q24 MDx User Manual 01/2016
5-13
3.
Switch on the vacuum pump.
4.
Apply vacuum to the tool by opening the vacuum switch.
5.
Wash the filter probes by lowering the probes into high-
purity water. Flush the probes with 70 ml high-purity
water.
Ensure that the water is being transferred to the waste
container. If it is not, ensure the tubing is connected
correctly and is not broken. Broken tubing should be
replaced (see Section 6.3.4).
Ensure that the waste filter is dry. If the filter is wet, it
should be replaced (see Section 6.3.5).
6.
Close the vacuum switch on the tool (Off) and place the
tool in the Parking position.
7.
Refill trough 5 with 70 ml high-purity water.
8.
Dilute the sequencing primer to 0.3 µM in Annealing
Buffer. Add 25 µl of the solution to each well of a
PyroMark Q24 Plate that is to be used.
Note
: Use one of the supplied PyroMark Q24 Plate
Holders as support when preparing and moving the
plate.