Maintenance
PyroMark Q24 MDx User Manual 01/2016
6-7
Procedure
1.
Ensure that no vacuum is applied to the vacuum prep
tool, i.e., the vacuum switch is closed (Off), and the
vacuum pump is switched off.
2.
Disconnect the vacuum pump from the mains power.
3.
Remove the tool from the tubing.
4.
Loosen the four screws using the 2 mm Allen key
supplied with the system.
5.
Pull out the old filter probes.
6.
Gently insert new filter probes without pressing on the
filter tips.
7.
Fasten the four screws and reconnect the vacuum pump
to the mains power.
6.3.3
Replacing the rubber seal
If the filter probes are loose and/or fall out, there are two
possible causes:
The four screws are not tight enough
The rubber seal needs to be replaced
If the rubber seal needs to be replaced, the following items
are required:
Powder-free gloves,
2 mm Allen key (supplied with the system)
New rubber seal (QIAGEN)