Operating Procedures
PyroMark Q24 MDx User Manual 01/2016
5-7
Note
: To base your run on a previous run, right-click the
processed run file in the shortcut browser and select “Copy
and Rerun” from the context menu. Only the run setup, not
the run and analysis data, will be copied.
5.2.4
Managing instrument methods
The instrument method should be selected according to the
reagents and reagent cartridge that will be used for the run.
The method number printed on the PyroMark Q24 Cartridge
corresponds to specific method settings provided at
www.qiagen.com/Products/PyroMarkQ24MDx.aspx.
Note
: It is recommended that only methods supplied by
QIAGEN are used.
To import a new method:
1.
From the Web site above, download the method file
corresponding to the method number printed on the
cartridge label. Save it to the computer running the
PyroMark Q24 MDx Software.
2.
In the “Instrument Methods” dialog box, click “Import”.
The “Find Instrument Method” dialog box opens.
3.
Locate and select the downloaded method and click
“Open”.
To create a new method:
1.
In the “Instrument method” dialog box, select an existing
method and click “Save As”.
2.
Enter a name for the new method and press “Enter”.
3.
Change the method settings in the dialog box to match
those posted on
www.qiagen.com/Products/PyroMarkQ24MDx.aspx.
4.
Click “Save”.
Summary of Contents for PyroMark Q24 MDx
Page 68: ...Maintenance 6 10 PyroMark Q24 MDx User Manual 01 2016 This page intentionally left blank ...
Page 76: ...Troubleshooting 7 8 PyroMark Q24 MDx User Manual 01 2016 This page left intentionally blank ...
Page 82: ...Glossary 8 6 PyroMark Q24 MDx User Manual 01 2016 This page intentionally left blank ...
Page 98: ...Appendix B B 16 PyroMark Q24 MDx User Manual 01 2016 This page intentionally left blank ...
Page 100: ...Appendix C C 2 PyroMark Q24 MDx User Manual 01 2016 This page intentionally left blank ...
Page 103: ......