Maintenance
6-8
PyroMark Q24 MDx User Manual 01/2016
Procedure
1.
Ensure that no vacuum is applied to the vacuum prep
tool, i.e., the vacuum switch is closed (Off), and the
vacuum pump is switched off.
2.
Disconnect the vacuum pump from the mains power.
3.
Remove the tool from the tubing.
4.
Remove the four screws using the 2 mm Allen key
supplied with the system.
5.
Gently remove the filter probes. Avoid contaminating the
filter probes.
6.
Remove the metal plate and replace the rubber seal.
7.
Reassemble the tool and reconnect the vacuum pump to
the mains power.
8.
Check that the filter probes are functioning properly by
performing the function test, as described in Section 5.3.1.
6.3.4
Replacing the tubing
If the tubing is broken or distorted, replace it.
Be sure to observe all national, state, and local
environmental regulations for the disposal of laboratory
waste.
The following items are required:
New tubing (QIAGEN)
Beaker
Procedure
1.
Ensure that no vacuum is applied to the vacuum prep
tool, i.e., the vacuum switch is closed (Off), and the
vacuum pump is switched off.
Summary of Contents for PyroMark Q24 MDx
Page 68: ...Maintenance 6 10 PyroMark Q24 MDx User Manual 01 2016 This page intentionally left blank ...
Page 76: ...Troubleshooting 7 8 PyroMark Q24 MDx User Manual 01 2016 This page left intentionally blank ...
Page 82: ...Glossary 8 6 PyroMark Q24 MDx User Manual 01 2016 This page intentionally left blank ...
Page 98: ...Appendix B B 16 PyroMark Q24 MDx User Manual 01 2016 This page intentionally left blank ...
Page 100: ...Appendix C C 2 PyroMark Q24 MDx User Manual 01 2016 This page intentionally left blank ...
Page 103: ......