Appendix B
PyroMark Q24 MDx User Manual 01/2016
B-1
Appendix A
Technical data
QIAGEN reserves the right to change specifications at any time.
Environmental conditions
Operating conditions: PyroMark Q24 MDx Instrument
Input power
100–240 VAC, 50–60 Hz
Input power
consumption
Maximum 160 VA
Instrument rating 24 V DC, 40 W
Cooler rating
12 V DC, 60 W
Overvoltage
category
II
Air temperature 15–32ºC (59–90ºF)
Relative humidity 20–90% (noncondensing)
Altitude
Up to 2000 m (6500 ft.)
Place of operation For indoor use only
Draft-free location, not close to window. Keep instrument
out of direct sunlight
Pollution level
2
Environmental
class
3K2 (IEC 60721-3-3)