Operating Procedures
5-12
PyroMark Q24 MDx User Manual 01/2016
The mains plug should be easily accessible in case the
vacuum pump needs to be disconnected quickly from the
mains power.
Note
: Perform the function test to ensure that the filter
probes are working properly (see Section 5.3.1). All filter
probes should be replaced after preparation of
approximately 100 plates.
Note
: Empty the waste container if necessary (see
Section 5.6.2).
2.
Fill five separate troughs supplied with the PyroMark
Q24 MDx Vacuum Workstation as follows:
Approximately 50 ml ethanol (70%) (1)
Approximately 40 ml Denaturation Solution (2)
Approximately 50 ml 1x Wash Buffer (3)
Approximately 50 ml high-purity water (4)
Approximately 70 ml high-purity water (5)
A suggested setup is shown in the following image. Refill
the troughs to these levels whenever necessary.