Installation Procedures
PyroMark Q24 MDx User Manual 01/2016
4-3
site. Mains supply voltage fluctuations are not to exceed 10%
of nominal supply voltages.
Grounding requirements
To protect operating personnel, the PyroMark Q24 MDx
Instrument must be correctly grounded (earthed). The
PyroMark Q24 MDx Instrument is equipped with two 3-
conductor AC power cords. To preserve this protection
feature, do not operate the PyroMark Q24 MDx Instrument
from AC power outlets that have no ground (earth)
connection.
4.3
Installation of the analysis software
4.3.1
Installing or upgrading PyroMark Q24 MDx
Software
Note
: If the computer is connected to a network, network
policy settings may prevent you from completing this
procedure. For more information, contact your system
administrator.
1.
Ensure that the computer meets the minimum
requirements; see Section 3.6.
2.
Close any programs running on the computer.
3.
Insert the PyroMark Q24 MDx Software CD into the CD drive.
4.
In the CD menu, click “Install PyroMark Q24 MDx
Software”.
If the CD menu does not appear automatically:
Select “(My) Computer” in the Windows “Start”
menu.
Right-click the CD drive with the software CD and
select “Open”.
Double-click the file
autorun.exe
.
5.
Follow the instructions that appear in the “Setup Wizard”.
Note
: If .NET Framework 3.5 has to be installed
(installation is prompted by the “Setup Wizard”), the
installation has to be restarted when the .NET Framework
installation has been completed, i.e., open the CD menu
Summary of Contents for PyroMark Q24 MDx
Page 68: ...Maintenance 6 10 PyroMark Q24 MDx User Manual 01 2016 This page intentionally left blank ...
Page 76: ...Troubleshooting 7 8 PyroMark Q24 MDx User Manual 01 2016 This page left intentionally blank ...
Page 82: ...Glossary 8 6 PyroMark Q24 MDx User Manual 01 2016 This page intentionally left blank ...
Page 98: ...Appendix B B 16 PyroMark Q24 MDx User Manual 01 2016 This page intentionally left blank ...
Page 100: ...Appendix C C 2 PyroMark Q24 MDx User Manual 01 2016 This page intentionally left blank ...
Page 103: ......