![Qiagen PyroMark Q24 MDx User Manual Download Page 23](http://html1.mh-extra.com/html/qiagen/pyromark-q24-mdx/pyromark-q24-mdx_user-manual_3179292023.webp)
General Description
PyroMark Q24 MDx User Manual 01/2016
3-5
3.4.1
Process chamber
Heating block.
The process chamber contains a heating block that
maintains the correct temperature of the plate and its
contents. If the room temperature is too high, the heating
block is cooled by the PyroMark Q24 MDx Instrument
Cooling Device (preinstalled).
Data are collected from all the wells simultaneously by 24
CCDs underneath the heating block. In wells where there is a
positive reaction with the added nucleotide light is emitted
giving rise to a peak in Pyrogram. To enable rapid mixing of
samples and reagents in the plate, the heating block inside
the process chamber is constantly vibrated during the run.
3.4.2
Dispensing unit
Dispensing unit.
A reagent cartridge (the PyroMark Q24 Cartridge) filled with
the required volumes of PyroMark Gold Q24 Reagents is
inserted into the dispensing unit. The instrument starts
Summary of Contents for PyroMark Q24 MDx