Safety Information
PyroMark Q24 MDx User Manual 01/2016
1-5
WARNING
Sharp needles
Do not touch the sharp needles at the bottom of the
reagent cartridge. Handle the needles with care. Small
particles and fibers may obstruct the needles.
CAUTION
Light guide maintenance
Use lint free tissues to clean the space between the heating
block and the light guide block inside the instrument. Do
not use paper tissues.
1.6
Heat hazard
WARNING
Hot surface
The plate holder and the heating block (for annealing) can
reach temperatures of up to 80°C (176°F). Avoid touching
them when they are hot.
CAUTION
Risk of overheating
To ensure proper ventilation, maintain a minimum
clearance of 10 cm (3.94 in.) at the sides and rear of the
PyroMark Q24 MDx Instrument.
Slits and openings that ensure the ventilation of the
PyroMark Q24 MDx Instrument must not be covered.
1.7
Consumables
CAUTION
Unsupported consumables
Do not connect or use any consumables, accessories, or
external equipment other than that specified.