Operating Procedures
PyroMark Q24 MDx User Manual 01/2016
5-21
Saving
The run data is transferred to the USB stick.
Do not remove the USB stick until the
instrument confirms that the run file has been
saved.
Finished
The run has been completed and the run data
have been transferred to the USB stick.
Pyrogram and warnings
The run name and selected well are displayed in the top left
corner. To select another well, use the
and
screen buttons.
Any instrument warnings are displayed below the Pyrogram
area (the three latest warnings are displayed). For suggested
actions, see Section 7.3.
Abort the run
To abort the run, press “Stop”.
5.5.5
After the run
1.
When the instrument confirms that the run file has been
saved to the USB stick, press “Close”.
2.
Remove the USB stick.
3.
Open the instrument lid.
4.
Open the cartridge gate and remove the reagent
cartridge by lifting it up and pulling it out.
5.
Close the gate.
6.
Open the plate-holding frame and remove the plate
from the heating block.
7.
Close the plate-holding frame and the instrument lid.
8.
Discard the plate.
9.
If the reagent cartridge is to be reused, clean it according
to the instructions below.
10.
If this was the last run for the day, follow the instructions
in Section 5.6.
Summary of Contents for PyroMark Q24 MDx
Page 68: ...Maintenance 6 10 PyroMark Q24 MDx User Manual 01 2016 This page intentionally left blank ...
Page 76: ...Troubleshooting 7 8 PyroMark Q24 MDx User Manual 01 2016 This page left intentionally blank ...
Page 82: ...Glossary 8 6 PyroMark Q24 MDx User Manual 01 2016 This page intentionally left blank ...
Page 98: ...Appendix B B 16 PyroMark Q24 MDx User Manual 01 2016 This page intentionally left blank ...
Page 100: ...Appendix C C 2 PyroMark Q24 MDx User Manual 01 2016 This page intentionally left blank ...
Page 103: ......