Draft - 1 Aug 08
Applying the Sensor
SpO
2
Monitoring
8-13
Inspect the application site every 2 to 3 hours or according to clinical practice
guidelines to ensure skin integrity and correct optical alignment. If skin
integrity changes, move the sensor to another site.
Follow the sensor’s instructions for use, adhering to all warnings and cautions.
To apply the sensor, use the following directions.
Adult Finger
sensor
(M1191A/B)
Push the sensor over the fingertip in such a way that the fingertip touches but
does not protrude from the end of the sensor. The fingertip must be uppermost
and the cable must lie on the back of the hand. This ensures that the light source
cover the base of the fingertip giving the best measurement results. The cable
can be held in place by the accompanying wristband.
Step
Action
1
Select the site and appropriate sensor (see “Selecting a SpO
2
Apply the sensor to the appropriate part of the patient’s body.
Note—
The application site should match the sensor size so that the
sensor can neither fall off nor apply excessive pressure.
3
Check that the light emitter and the photo detector are directly
opposite each other. All light from the emitter must pass through the
patient’s tissue.
Summary of Contents for IntelliVue TRx
Page 6: ...Draft 1 Aug 08 About this Book vi ...
Page 40: ...Draft 1 Aug 08 Accessories 2 18 Product Safety ...
Page 54: ...Draft 1 Aug 08 Audible Tones 3 14 Transceiver Controls ...
Page 74: ...Draft 1 Aug 08 Battery Information 4 20 Basic Operation ...
Page 94: ...Draft 1 Aug 08 Technical Alarms INOPs 5 20 Alarms ...
Page 152: ...Draft 1 Aug 08 How the QT Analysis Algorithm Works 7 24 ST AR Arrhythmia Monitoring ...
Page 176: ...Draft 1 Aug 08 Optimizing SpO2 Measurement Performance 8 24 SpO2 Monitoring ...
Page 236: ...Draft 1 Aug 08 Troubleshooting 11 16 Maintenance Cleaning Troubleshooting ...
Page 268: ...Draft 1 Aug 08 SpO2 Accessories A 10 Accessories ...
Page 270: ...Draft 1 Aug 08 B 2 Sales and Support Offices ...
Page 274: ...Draft 1 Aug 08 Index 4 ...