Applying the Sensor
8-14
SpO
2
Monitoring
Small Adult/
Pediatric
Finger sensor
(M1192A)
Push the sensor over the fingertip in such a way that the fingertip touches but
does not protrude from the end of the sensor.
Ear Clip sensor
(M1194A)
Clip the probe onto the fleshy part of the ear lobe as shown in the diagram
below. The plastic fixing mechanism helps to minimize artifact generated by
patient motion. Do not position the probe on cartilage or where it presses
against the head.
The clip sensor can be used as an alternative if the adult finger sensor does not
provide satisfactory results. The preferred application site is the ear lobe,
although other application sites with higher perfusion (such as the nostril) can
be used. Due to the physiologically lower perfusion in the ear lobe, you should
be aware of the reduced accuracy of the measurement and more frequent
technical alarms.
Summary of Contents for IntelliVue TRx
Page 6: ...Draft 1 Aug 08 About this Book vi ...
Page 40: ...Draft 1 Aug 08 Accessories 2 18 Product Safety ...
Page 54: ...Draft 1 Aug 08 Audible Tones 3 14 Transceiver Controls ...
Page 74: ...Draft 1 Aug 08 Battery Information 4 20 Basic Operation ...
Page 94: ...Draft 1 Aug 08 Technical Alarms INOPs 5 20 Alarms ...
Page 152: ...Draft 1 Aug 08 How the QT Analysis Algorithm Works 7 24 ST AR Arrhythmia Monitoring ...
Page 176: ...Draft 1 Aug 08 Optimizing SpO2 Measurement Performance 8 24 SpO2 Monitoring ...
Page 236: ...Draft 1 Aug 08 Troubleshooting 11 16 Maintenance Cleaning Troubleshooting ...
Page 268: ...Draft 1 Aug 08 SpO2 Accessories A 10 Accessories ...
Page 270: ...Draft 1 Aug 08 B 2 Sales and Support Offices ...
Page 274: ...Draft 1 Aug 08 Index 4 ...