Physiologic Alarms
5-4
Alarms
Unsuspending
& Resuming
Alarms
Alarms will be resumed automatically after the configured suspend duration
time. You can cancel alarm suspend manually before the configured suspend
duration time has expired from the transceiver (see following directions) or from
the Information Center.
Physiologic Alarms
Physiologic alarms indicate a life-threatening situation or a less urgent situation
such as heart rate beyond limits. There are no physiologic alarm signals
generated by the transceiver.
All physiologic alarms are generated at the IntelliVue Information Center,
and all alarm signals must be acknowledged at the Information Center.
Step
Action
1
Press the
Telemetry
and
Check
buttons simultaneously until the
indicator turns off.
•
The Alarm Suspend icon is turned off.
•
The message "Alarms Suspended" (or "Alarms Paused") is
removed from the Information Center display.
2
Alternately, you can unsuspend alarms at the Information Center.
See
IntelliVue Information Center Instructions for Use
or
Online
Hel
p for directions.
For instructions on how to resume alarms from TeleMon, see the
TeleMon Instructions for Use.
Summary of Contents for IntelliVue TRx
Page 6: ...Draft 1 Aug 08 About this Book vi ...
Page 40: ...Draft 1 Aug 08 Accessories 2 18 Product Safety ...
Page 54: ...Draft 1 Aug 08 Audible Tones 3 14 Transceiver Controls ...
Page 74: ...Draft 1 Aug 08 Battery Information 4 20 Basic Operation ...
Page 94: ...Draft 1 Aug 08 Technical Alarms INOPs 5 20 Alarms ...
Page 152: ...Draft 1 Aug 08 How the QT Analysis Algorithm Works 7 24 ST AR Arrhythmia Monitoring ...
Page 176: ...Draft 1 Aug 08 Optimizing SpO2 Measurement Performance 8 24 SpO2 Monitoring ...
Page 236: ...Draft 1 Aug 08 Troubleshooting 11 16 Maintenance Cleaning Troubleshooting ...
Page 268: ...Draft 1 Aug 08 SpO2 Accessories A 10 Accessories ...
Page 270: ...Draft 1 Aug 08 B 2 Sales and Support Offices ...
Page 274: ...Draft 1 Aug 08 Index 4 ...