Draft - 1 Aug 08
Optimizing SpO
2
Measurement Performance
SpO
2
Monitoring
8-23
Care and Cleaning
For care and cleaning instructions, see the instructions accompanying the
sensors.
Sensor Wear
Normal wear and tear associated with patient movement and regular sensor
cleaning typically mean that the sensor will have a limited lifetime. However,
provided you handle the sensor and its cable with care, you can expect useful
service from it for up to two years. Harsh treatment will drastically reduce the
lifetime of the sensor. Moreover, Philips Medical Systems’ warranty agreement
shall not apply to defects arising from improper use.
To get the best results from your SpO
2
reusable sensor:
•
Always handle the sensor and cable with care. The soft finger sleeve
houses a sensitive electronic device that can be damaged by harsh
treatment. Always protect the cable from sharp-edged objects.
•
Use the wristband that is supplied with your M1191A sensor. By keeping
the cable between the finger sensor and the wristband fairly loose, you
will maintain good monitoring conditions.
Summary of Contents for IntelliVue TRx
Page 6: ...Draft 1 Aug 08 About this Book vi ...
Page 40: ...Draft 1 Aug 08 Accessories 2 18 Product Safety ...
Page 54: ...Draft 1 Aug 08 Audible Tones 3 14 Transceiver Controls ...
Page 74: ...Draft 1 Aug 08 Battery Information 4 20 Basic Operation ...
Page 94: ...Draft 1 Aug 08 Technical Alarms INOPs 5 20 Alarms ...
Page 152: ...Draft 1 Aug 08 How the QT Analysis Algorithm Works 7 24 ST AR Arrhythmia Monitoring ...
Page 176: ...Draft 1 Aug 08 Optimizing SpO2 Measurement Performance 8 24 SpO2 Monitoring ...
Page 236: ...Draft 1 Aug 08 Troubleshooting 11 16 Maintenance Cleaning Troubleshooting ...
Page 268: ...Draft 1 Aug 08 SpO2 Accessories A 10 Accessories ...
Page 270: ...Draft 1 Aug 08 B 2 Sales and Support Offices ...
Page 274: ...Draft 1 Aug 08 Index 4 ...