Draft - 1 Aug 08
How the QT Analysis Algorithm Works
ST/AR Arrhythmia Monitoring
7-19
Adjustment
Description
Turn QT Analysis On/Off
Turn QT analysis on by clicking in the
QT
Analysis On
check box. QT analysis is on
when a checkmark displays in the check box.
When the QT measurement is on, a QT status
message is displayed in the QT Setup window,
along with the current values for QT, QTc, dQTc
and QT-HR. The lead labels indicating the leads
used to calculate the baseline and current values
also appear.
Note—
Turning QT analysis off does not clear
the baseline value. This allows you to turn QT
analysis off during prolonged arrhythmias, such
as bigeminy, without losing the baseline.