Draft - 1 Aug 08
Technical Alarms (INOPs)
5-14
Alarms
MORE BED ALARMS
Source - Patient Monitor
Red or
Yellow or
Hard
Technical
Alarm
The monitor is associated
with a transceiver and is
sending data to the
Information Center via the
transceiver. There are
currently more alarms at
the bedside than can be
transmitted to the
Information Center.
For more information only.
Appears at Information
Center only.
NO ARRHYTHMIA
Note—
This condition
displays as All ARRH
ALRMS OFF at the IIC.
Source - Information
Center
Soft
All basic ST/AR ECG
alarms have been turned
off.
Turn arrhythmia alarms on.
NO ECG AT CENTRAL
Source - Patient Monitor
Hard
The ECG measured with
the monitor is not being
sent to the Information
Center via the transceiver.
Check equipment
connections and
configuration settings.
NO ECG SOURCE
Source - Telemetry
Hard
The transceiver is paired
with a monitor but the
Information Center is not
detecting a valid ECG
signal from either of them.
Check equipment
connections and
configuration settings.
Technical Alarms (Inoperative Conditions)
Alarm Text
Priority
Condition
What to do
Summary of Contents for IntelliVue TRx
Page 6: ...Draft 1 Aug 08 About this Book vi ...
Page 40: ...Draft 1 Aug 08 Accessories 2 18 Product Safety ...
Page 54: ...Draft 1 Aug 08 Audible Tones 3 14 Transceiver Controls ...
Page 74: ...Draft 1 Aug 08 Battery Information 4 20 Basic Operation ...
Page 94: ...Draft 1 Aug 08 Technical Alarms INOPs 5 20 Alarms ...
Page 152: ...Draft 1 Aug 08 How the QT Analysis Algorithm Works 7 24 ST AR Arrhythmia Monitoring ...
Page 176: ...Draft 1 Aug 08 Optimizing SpO2 Measurement Performance 8 24 SpO2 Monitoring ...
Page 236: ...Draft 1 Aug 08 Troubleshooting 11 16 Maintenance Cleaning Troubleshooting ...
Page 268: ...Draft 1 Aug 08 SpO2 Accessories A 10 Accessories ...
Page 270: ...Draft 1 Aug 08 B 2 Sales and Support Offices ...
Page 274: ...Draft 1 Aug 08 Index 4 ...