Technical Alarms (INOPs)
Alarms
5-11
In the following table, technical alarms are listed alphabetically.
Technical Alarms (Inoperative Conditions)
Alarm Text
Priority
Condition
What to do
BATTERY LOW
BATTERY LOW T
Source - Telemetry
Soft
Battery power is low.
There is less than 15 minutes
of monitoring time
remaining. Replace batteries
promptly to avoid transceiver
shutdown and cessation of
monitoring.
CANNOT ANALYZE
ECG
Source - Information
Center
Hard
Arrhythmia algorithm
cannot reliably analyze the
ECG data on any
monitored leads.
Assess the lead selections,
initiate relearn, and validate
analyzed rhythm. Check
other INOPs for possible
source of problem.
CANNOT ANALYZE QT
Source - Information
Center
Soft
There have been two
consecutive invalid 5
minute values.
See “Limitations for QT
Monitoring” on page 7-23.
CANNOT ANALYZE ST
Source - Information
Center
Soft
ST algorithm cannot
reliably generate any valid
ST values on any
monitored lead.
Review the ECG signal
quality and correct if
necessary. Reposition the
ISO and J points.
Note—
If the patient has a
ventricular pacemaker, ST
analysis is not possible.
Summary of Contents for IntelliVue TRx
Page 6: ...Draft 1 Aug 08 About this Book vi ...
Page 40: ...Draft 1 Aug 08 Accessories 2 18 Product Safety ...
Page 54: ...Draft 1 Aug 08 Audible Tones 3 14 Transceiver Controls ...
Page 74: ...Draft 1 Aug 08 Battery Information 4 20 Basic Operation ...
Page 94: ...Draft 1 Aug 08 Technical Alarms INOPs 5 20 Alarms ...
Page 152: ...Draft 1 Aug 08 How the QT Analysis Algorithm Works 7 24 ST AR Arrhythmia Monitoring ...
Page 176: ...Draft 1 Aug 08 Optimizing SpO2 Measurement Performance 8 24 SpO2 Monitoring ...
Page 236: ...Draft 1 Aug 08 Troubleshooting 11 16 Maintenance Cleaning Troubleshooting ...
Page 268: ...Draft 1 Aug 08 SpO2 Accessories A 10 Accessories ...
Page 270: ...Draft 1 Aug 08 B 2 Sales and Support Offices ...
Page 274: ...Draft 1 Aug 08 Index 4 ...