Draft - 1 Aug 08
Technical Alarms (INOPs)
5-18
Alarms
TELE CONFIG
UNSUPPORTED
Source - Patient Monitor
Hard
Transceiver not supported
(companion mode).
Make sure the transceiver is
the correct revision.
Contact Service.
TELE DISCONNECTED
Sourece - Telemetry
Red,
Yellow or
Hard
Technical
Alarm
•
The interface cable is
not connected.
•
Short-range radio
connection between
the transceiver and
the patient monitor
has been lost.
If the disconnection is not
intentional:
•
Re-connect the
interface cable.
•
Identify and remove
interference sources
such as bluetooth
devices, DECT
phones, microwaves,
etc.
•
Reduce the number of
devices equipped with
short-range radio
capability.
TELE INCOMPATIBLE
Source - Patient Monitor
Hard
Transceivers equipped with
SRRA are not supported
with this revision of the
Information Center.
Check equipment
configuration/compatibility.
Contact Service.
TELEMETRY STANDBY
Soft
Information Center standby
mode timer is active, or
patient was not returned to
telemetry coverage area.
There is no data from bed.
Cancelled when patient is
removed from Standby.
Technical Alarms (Inoperative Conditions)
Alarm Text
Priority
Condition
What to do
Summary of Contents for IntelliVue TRx
Page 6: ...Draft 1 Aug 08 About this Book vi ...
Page 40: ...Draft 1 Aug 08 Accessories 2 18 Product Safety ...
Page 54: ...Draft 1 Aug 08 Audible Tones 3 14 Transceiver Controls ...
Page 74: ...Draft 1 Aug 08 Battery Information 4 20 Basic Operation ...
Page 94: ...Draft 1 Aug 08 Technical Alarms INOPs 5 20 Alarms ...
Page 152: ...Draft 1 Aug 08 How the QT Analysis Algorithm Works 7 24 ST AR Arrhythmia Monitoring ...
Page 176: ...Draft 1 Aug 08 Optimizing SpO2 Measurement Performance 8 24 SpO2 Monitoring ...
Page 236: ...Draft 1 Aug 08 Troubleshooting 11 16 Maintenance Cleaning Troubleshooting ...
Page 268: ...Draft 1 Aug 08 SpO2 Accessories A 10 Accessories ...
Page 270: ...Draft 1 Aug 08 B 2 Sales and Support Offices ...
Page 274: ...Draft 1 Aug 08 Index 4 ...