Draft - 1 Aug 08
Measuring ECG
6-6
ECG Monitoring
There is no cardiotach within the transceiver; cardiotach analysis resides in the
arrhythmia algorithm at the Information Center. Therefore, arrhythmia analysis
is always turned on for telemetry patients. Arrhythmia analysis is either Basic or
Enhanced, depending on the product configuration.
ECG Leads
Monitored
Depending on the leadset connected to the transceiver, a different set of
viewable leads are available at the Information Center. The transceiver can
source up to four raw ECG waves.
If you are using ...
these leads can be selected at the
Information Center
3-wire
I, II, III
Sourced (raw) waves are received as:
•
Channel 1 = I, II, or III
Default is II.
5-wire (Standard mode)
I, II, III, aVR, aVL, aVF, MCL and V
Sourced (raw) waves are received as:
•
Channel 1 = II
•
Channel 2 = III
•
Channel 3 = MCL
Defaults are II, V, III.
Summary of Contents for IntelliVue TRx
Page 6: ...Draft 1 Aug 08 About this Book vi ...
Page 40: ...Draft 1 Aug 08 Accessories 2 18 Product Safety ...
Page 54: ...Draft 1 Aug 08 Audible Tones 3 14 Transceiver Controls ...
Page 74: ...Draft 1 Aug 08 Battery Information 4 20 Basic Operation ...
Page 94: ...Draft 1 Aug 08 Technical Alarms INOPs 5 20 Alarms ...
Page 152: ...Draft 1 Aug 08 How the QT Analysis Algorithm Works 7 24 ST AR Arrhythmia Monitoring ...
Page 176: ...Draft 1 Aug 08 Optimizing SpO2 Measurement Performance 8 24 SpO2 Monitoring ...
Page 236: ...Draft 1 Aug 08 Troubleshooting 11 16 Maintenance Cleaning Troubleshooting ...
Page 268: ...Draft 1 Aug 08 SpO2 Accessories A 10 Accessories ...
Page 270: ...Draft 1 Aug 08 B 2 Sales and Support Offices ...
Page 274: ...Draft 1 Aug 08 Index 4 ...