Turning the Transceiver On/Off
4-4
Basic Operation
RF Auto
Shutoff
When RF Auto Shutoff is enabled (default), the transceiver stops broadcasting a
radio signal in order to prevent interference with other transceivers in use. This
condition occurs if there is no ECG signal for 10 minutes and the SpO
2
sensor
cable is
not
inserted in the SpO
2
sensor port. The technical alarm, "No Signal",
followed by the "Transmitter Off" INOP will be displayed at the Information
Center
.
Remove the batteries to conserve battery charge. However, if there is no
ECG signal for 10 minutes but the SpO
2
cable
is
inserted, the transceiver does
not stop broadcasting the signal.
To restart monitoring, insert batteries (if needed), attach leads to the patient,
insert the SpO
2
sensor cable in the SpO
2
sensor port and press the Check button
to verify association with the Information Center.
Standby
Mode
Standby mode is used to temporarily suspend monitoring at the Information
Center, for example, when a patient goes out of access point range or is
discharged.
Standby does not power down the transceiver, so if the standby period is
prolonged, you should remove the batteries.
Initiating
Standby
Step
Action
1
Do not remove the leads until after the patient is in Standby. This
will avoid an "ECG Leads Off" technical alarm, as well as
reminders, if configured.
2
In the Patient Window, select the
Standby
button.
3
Select the duration of the standby period (depending on your
system’s configuration: Infinite, 4 Hours, 3 Hours, 2 Hours, 1 Hour,
30 Minutes, 20 Minutes, or 10 Minutes).
Note—
If the patient will be discharged, select “Infinite” as the
standby duration.
Summary of Contents for IntelliVue TRx
Page 6: ...Draft 1 Aug 08 About this Book vi ...
Page 40: ...Draft 1 Aug 08 Accessories 2 18 Product Safety ...
Page 54: ...Draft 1 Aug 08 Audible Tones 3 14 Transceiver Controls ...
Page 74: ...Draft 1 Aug 08 Battery Information 4 20 Basic Operation ...
Page 94: ...Draft 1 Aug 08 Technical Alarms INOPs 5 20 Alarms ...
Page 152: ...Draft 1 Aug 08 How the QT Analysis Algorithm Works 7 24 ST AR Arrhythmia Monitoring ...
Page 176: ...Draft 1 Aug 08 Optimizing SpO2 Measurement Performance 8 24 SpO2 Monitoring ...
Page 236: ...Draft 1 Aug 08 Troubleshooting 11 16 Maintenance Cleaning Troubleshooting ...
Page 268: ...Draft 1 Aug 08 SpO2 Accessories A 10 Accessories ...
Page 270: ...Draft 1 Aug 08 B 2 Sales and Support Offices ...
Page 274: ...Draft 1 Aug 08 Index 4 ...