Technical Alarms (INOPs)
Alarms
5-15
NO SIGNAL
Source - Information
Center
Hard
•
Transceiver is
outside the coverage
area, or
•
No batteries in
transceiver, or
•
Transceiver has
failed, or
•
Connection to
Database Server has
failed.
•
Short-range radio
connection has
failed.
•
Make sure that the
transceiver is within
the coverage area and
has good batteries.
•
Replace the transceiver
if Power On Self Test
fails.
•
Put bed in Standby.
•
Contact Service.
OUT OF AREA
Source - Information
Center
Hard
The transceiver is
communicating with an
access point designated as a
boundary limit.
For more information only.
REPLACE BATTERY
REPLACE BATTERY T
Note—
This INOP may also
be configured to display as
a Red or Yellow Technical
Alarm.
Source - Telemetry
Red or
Yellow or
Hard
Technical
Alarm,
Latched
Dead battery. No
monitoring is occurring.
Replace batteries.
SOME ECG ALRMS OFF
Source - Information
Center
Soft
Some yellow arrhythmia
alarms have been turned off
for this patient.
For information only.
Technical Alarms (Inoperative Conditions)
Alarm Text
Priority
Condition
What to do
Summary of Contents for IntelliVue TRx
Page 6: ...Draft 1 Aug 08 About this Book vi ...
Page 40: ...Draft 1 Aug 08 Accessories 2 18 Product Safety ...
Page 54: ...Draft 1 Aug 08 Audible Tones 3 14 Transceiver Controls ...
Page 74: ...Draft 1 Aug 08 Battery Information 4 20 Basic Operation ...
Page 94: ...Draft 1 Aug 08 Technical Alarms INOPs 5 20 Alarms ...
Page 152: ...Draft 1 Aug 08 How the QT Analysis Algorithm Works 7 24 ST AR Arrhythmia Monitoring ...
Page 176: ...Draft 1 Aug 08 Optimizing SpO2 Measurement Performance 8 24 SpO2 Monitoring ...
Page 236: ...Draft 1 Aug 08 Troubleshooting 11 16 Maintenance Cleaning Troubleshooting ...
Page 268: ...Draft 1 Aug 08 SpO2 Accessories A 10 Accessories ...
Page 270: ...Draft 1 Aug 08 B 2 Sales and Support Offices ...
Page 274: ...Draft 1 Aug 08 Index 4 ...