Draft - 1 Aug 08
ST/AR Arrhythmia Algorithm
7-6
ST/AR Arrhythmia Monitoring
Beat classification determined by the ST/AR algorithm is shown on the primary
delayed wave in the Arrhythmia Analysis window at the Information Center. To
access this window, select
Arrhythmia Analysis
from the Patient Window.
The annotation requires clinical validation of the analyzed heart rhythm. If the
analysis is inaccurate, perform a relearn of the rhythm.
When monitoring is initiated, when the Primary Wave is changed, or if Relearn
is selected, a question mark (?) is displayed next to HR and the annotation “L”
appears on the annotated wave until the HR is calculated and the rhythm is
learned.
Annotation
Beat Classification
Display
Color
A
Artifact
Blue
I
Inoperative
Red
L
Learning
Red
M
Missed Beat
Red
N
Normal
Blue
P
Paced
Blue
S
Supraventricular Premature
Blue
V
Ventricular Premature
Red
?
Questionable
Red
'
Pacer Spike
Blue
"
Biventricular Pacer Spike
Blue
Summary of Contents for IntelliVue TRx
Page 6: ...Draft 1 Aug 08 About this Book vi ...
Page 40: ...Draft 1 Aug 08 Accessories 2 18 Product Safety ...
Page 54: ...Draft 1 Aug 08 Audible Tones 3 14 Transceiver Controls ...
Page 74: ...Draft 1 Aug 08 Battery Information 4 20 Basic Operation ...
Page 94: ...Draft 1 Aug 08 Technical Alarms INOPs 5 20 Alarms ...
Page 152: ...Draft 1 Aug 08 How the QT Analysis Algorithm Works 7 24 ST AR Arrhythmia Monitoring ...
Page 176: ...Draft 1 Aug 08 Optimizing SpO2 Measurement Performance 8 24 SpO2 Monitoring ...
Page 236: ...Draft 1 Aug 08 Troubleshooting 11 16 Maintenance Cleaning Troubleshooting ...
Page 268: ...Draft 1 Aug 08 SpO2 Accessories A 10 Accessories ...
Page 270: ...Draft 1 Aug 08 B 2 Sales and Support Offices ...
Page 274: ...Draft 1 Aug 08 Index 4 ...