Measuring ECG
ECG Monitoring
6-9
Reconstructed
Leads
Reconstruction of leads from the sourced wave is defined by the calculations in
the following table. EASI reconstructed leads are a linear combination of all
three raw EASI leads.
ECG Lead
Clinical Calculations
in terms of electrodes
3-Wire
5-Wire
Standard
6-Wire
I
I
I
LA-RA
II
(default)
II
(default)
II
(default)
LL-RA
III
III
(default)
III
(default)
LL-LA
-
MCL
V-LA, where V=C
-
aVR
aVR
RA-(LA+LL)/2
-
aVL
aVL
LA-(RA+LL)/2
-
aVF
aVF
LL-(LA+RA)/2
-
V
(default)
V-(RA+LA+LL)/3, where V=C
Va
Va-(RA+LA+LL)/3, where
Va=V2 (default) position
Vb
Vb-(RA+LA+LL)/3, where Vb
=V5 (default) position