Draft - 1 Aug 08
Pairing Networked Devices
Pairing Monitoring Devices
10-9
Where Initiated
Effect at Bedside
Effect at Information Center
Silence
Bedside
Silences bedside alarms.
Silences telemetry alarms if Enable
Global Silence for Paired Beds is on at
the Information Center.
Silences bedside alarms.
Silences telemetry alarms if Enable
Global Silence for Paired Beds is on
at the Information Center.
Bedside from
Telemetry Data
Window
Silences bedside and telemetry alarms.
Silence Overview Alarms at Bedside
must be enabled at the Information
Center.
Silences bedside and telemetry
alarms.
Silence Overview Alarms at Bedside
must be enabled at the Information
Center.
Information Center
Silences bedside and telemetry alarms.
Silences bedside and telemetry
alarms.