Draft - 1 Aug 08
Pairing Non-networked Devices
10-22
Pairing Monitoring Devices
Short-
Range
Radio Error
Conditions
The following table describes error conditions that may occur when the
transceiver is paired with the monitor (MP5, MP2 and X2 IntelliVue Patient
Monitors only) via short-range radio capability.
Condition
Description
Action
Tele Disconnected INOP
displayed at the monitor
and the Information Center.
Short-range radio connection
between the transceiver and
MP5T has been lost due to a
failure of the short-range radio
connection.
There are too many short-range
radios operating in the same
vicinity (maximum of 4 per radio
channel).
If the disconnection is not
intentional:
Identify and remove the
interference sources.
Reduce the number of
devices equipped with
short-range radio capability.
No Signal INOP displayed
at the Information Center.
The short-range radio is not
installed correctly.
There is interference from
telemetry or ISM radio sources.
The transceiver is outside the
coverage area.
Note—
The typical coverage area
between the monitor and
transceiver is 15 ft.
Contact service personnel.
Identify and remove the
interference source.
Make sure the location of
the transceiver is in the
coverage area.
Battery LowT INOP
displayed at the monitor
and the Information Center.
When operating wirelessly, the
bedside monitor is no longer
providing power to the
transmitter, and battery capacity
is diminished. 15 minutes of
monitoring time remain.
Insert new AA batteries in
the transceiver.
Summary of Contents for IntelliVue TRx
Page 6: ...Draft 1 Aug 08 About this Book vi ...
Page 40: ...Draft 1 Aug 08 Accessories 2 18 Product Safety ...
Page 54: ...Draft 1 Aug 08 Audible Tones 3 14 Transceiver Controls ...
Page 74: ...Draft 1 Aug 08 Battery Information 4 20 Basic Operation ...
Page 94: ...Draft 1 Aug 08 Technical Alarms INOPs 5 20 Alarms ...
Page 152: ...Draft 1 Aug 08 How the QT Analysis Algorithm Works 7 24 ST AR Arrhythmia Monitoring ...
Page 176: ...Draft 1 Aug 08 Optimizing SpO2 Measurement Performance 8 24 SpO2 Monitoring ...
Page 236: ...Draft 1 Aug 08 Troubleshooting 11 16 Maintenance Cleaning Troubleshooting ...
Page 268: ...Draft 1 Aug 08 SpO2 Accessories A 10 Accessories ...
Page 270: ...Draft 1 Aug 08 B 2 Sales and Support Offices ...
Page 274: ...Draft 1 Aug 08 Index 4 ...