Cleaning and Sterilization
11-12
Maintenance, Cleaning & Troubleshooting
Test Procedure
Caution
You must perform this test each time you put a transceiver through the EO
sterilization process.
This test allows you to verify that patient information for both ECG and SpO
2
(if
you are monitoring pulse oximetry) appear at the Information Center and at the
bedside. You can use this procedure with a Patient Simulator.
Note—
This test assumes that the telemetry system and Information Center are
fully installed, and that you have performed the procedure to learn the
transceiver identity code.
Test the transceiver by performing the following steps. If the test indications do
not appear, refer to your service provider.
Step
Action
1
Perform a mechanical inspection of the transceiver (connectors,
battery door opening and closing, Telemetry and Check buttons).
2
At the Information Center, select the telemetry bedside you are
testing.
Summary of Contents for IntelliVue TRx
Page 6: ...Draft 1 Aug 08 About this Book vi ...
Page 40: ...Draft 1 Aug 08 Accessories 2 18 Product Safety ...
Page 54: ...Draft 1 Aug 08 Audible Tones 3 14 Transceiver Controls ...
Page 74: ...Draft 1 Aug 08 Battery Information 4 20 Basic Operation ...
Page 94: ...Draft 1 Aug 08 Technical Alarms INOPs 5 20 Alarms ...
Page 152: ...Draft 1 Aug 08 How the QT Analysis Algorithm Works 7 24 ST AR Arrhythmia Monitoring ...
Page 176: ...Draft 1 Aug 08 Optimizing SpO2 Measurement Performance 8 24 SpO2 Monitoring ...
Page 236: ...Draft 1 Aug 08 Troubleshooting 11 16 Maintenance Cleaning Troubleshooting ...
Page 268: ...Draft 1 Aug 08 SpO2 Accessories A 10 Accessories ...
Page 270: ...Draft 1 Aug 08 B 2 Sales and Support Offices ...
Page 274: ...Draft 1 Aug 08 Index 4 ...