Draft - 1 Aug 08
ST/AR ST Segment Algorithm
7-10
ST/AR Arrhythmia Monitoring
•
You are unable to get any lead that is not noisy.
•
Arrhythmias such as atrial fib/flutter cause irregular baseline.
•
The patient is continuously ventricularly paced.
•
The patient has left bundle branch block.
Adjusting ST
Measurement
Points
The ST Setup Window enables you to adjust the ST measurement points to
ensure accurate data.
There are three measurement cursors:
•
The ISO measurement cursor positions the isoelectric point in relation to
the R-wave peak.
•
The J-point cursor positions the J-point in relation to the R-wave peak.
The purpose of the J-point is to correctly position the ST measurement
point.
•
The ST measurement cursor positions the ST point a fixed distance from
the J point.
Note—
The ST measurement points may need to be re-adjusted if the patient's
heart rate or ECG morphology changes significantly.
Perform the following steps at the Information Center to adjust the ST
measurement points:
Step
Action
1
Access the ST Setup window by clicking on the
All Controls
button
in the Patient Window, then clicking on the
ST Setup
button.
2
If you need to adjust the ISO (isoelectric) point, place the cursor
over the
ISO
button to access the adjustment arrows. Then use the
arrows to position the bar in the middle of the flattest part of the
baseline (between the P and Q waves or in front of the P wave).
T
R
P
ISO
point
Q
S
Summary of Contents for IntelliVue TRx
Page 6: ...Draft 1 Aug 08 About this Book vi ...
Page 40: ...Draft 1 Aug 08 Accessories 2 18 Product Safety ...
Page 54: ...Draft 1 Aug 08 Audible Tones 3 14 Transceiver Controls ...
Page 74: ...Draft 1 Aug 08 Battery Information 4 20 Basic Operation ...
Page 94: ...Draft 1 Aug 08 Technical Alarms INOPs 5 20 Alarms ...
Page 152: ...Draft 1 Aug 08 How the QT Analysis Algorithm Works 7 24 ST AR Arrhythmia Monitoring ...
Page 176: ...Draft 1 Aug 08 Optimizing SpO2 Measurement Performance 8 24 SpO2 Monitoring ...
Page 236: ...Draft 1 Aug 08 Troubleshooting 11 16 Maintenance Cleaning Troubleshooting ...
Page 268: ...Draft 1 Aug 08 SpO2 Accessories A 10 Accessories ...
Page 270: ...Draft 1 Aug 08 B 2 Sales and Support Offices ...
Page 274: ...Draft 1 Aug 08 Index 4 ...