Physiologic Alarms
Alarms
5-7
** SpO
2
T yyy > xxx
Yellow
High SpO
2
.
SpO
2
value (yyy) greater than high SpO
2
limit (xxx).
SpO
2
** SpO
2
T yyy < xxxx
Yellow
Low SpO
2
.
SpO
2
value (yyy) less than low SpO
2
limit
(xxx).
SpO
2
** ST lead > xxx
Yellow
STlead is greater than the Elevation limit.
ST/AR
** ST lead < xxx
Yellow
STlead is less than the Depression limit.
ST/AR
* HR yyy > xxx
Yellow
Heart Rate (yyy) greater than the upper
Heart rate limit (xxx).
ST/AR
Basic &
Enhanced
Arrhythmia
* HR yyy < xxx
Yellow
Heart Rate (yyy) lower than the lower
Heart Rate limit (xxx).
ST/AR
Basic &
Enhanced
Arrhythmia
* IRREGULAR HR
Yellow
Consistently irregular rhythm (irregular
R-R intervals).
ST/AR
Enhanced
Arrhythmia
* MISSED BEAT
Yellow
No beat detected for 1.75 x average R-R
interval for Heart Rate greater than 120,
or no beat for 1 second with Heart Rate
greater than 120 (non-paced patient only).
ST/AR
Enhanced
Arrhythmia
* MULTIFORM PVCs
Yellow
The occurrence of two differently shaped
Vs, each occurring at least twice within
the last 300 beats as well as each
occurring at least once within the last 60
beats.
ST/AR
Enhanced
Arrhythmia
Physiologic (Patient) Alarms
Alarm Text
Priority
Condition
Source
Summary of Contents for IntelliVue TRx
Page 6: ...Draft 1 Aug 08 About this Book vi ...
Page 40: ...Draft 1 Aug 08 Accessories 2 18 Product Safety ...
Page 54: ...Draft 1 Aug 08 Audible Tones 3 14 Transceiver Controls ...
Page 74: ...Draft 1 Aug 08 Battery Information 4 20 Basic Operation ...
Page 94: ...Draft 1 Aug 08 Technical Alarms INOPs 5 20 Alarms ...
Page 152: ...Draft 1 Aug 08 How the QT Analysis Algorithm Works 7 24 ST AR Arrhythmia Monitoring ...
Page 176: ...Draft 1 Aug 08 Optimizing SpO2 Measurement Performance 8 24 SpO2 Monitoring ...
Page 236: ...Draft 1 Aug 08 Troubleshooting 11 16 Maintenance Cleaning Troubleshooting ...
Page 268: ...Draft 1 Aug 08 SpO2 Accessories A 10 Accessories ...
Page 270: ...Draft 1 Aug 08 B 2 Sales and Support Offices ...
Page 274: ...Draft 1 Aug 08 Index 4 ...