Draft - 1 Aug 08
Measurement Specifications
Safety Standards & Specifications
12-17
Gain Accuracy
+/- 5% at 25
o
C (77
o
F)
Noise Referred to ECG
Input (Peak-to-Peak)
AAMI: 30
μ
V (as per AAMI EC 13)
Lead Wires
3, 5 or 6-wire leadset compatible with IntelliVue
Patient Monitor, AAMI/IEC color codes
Time to baseline from
Defibrillator
AAMI: 5 s max (until ECG wave is on display
but not yet centered, monitoring bandwidth)
Pacer Rejection
Performance
(Pace pulses with no
tails).
Positive pacers
1
Amplitude Width
+2 to +700 mV 0.1, 0.2, 0.5 and 1.0 ms
+2 to +500 mV 1.5 ms
+2 to +400 mV 2 ms
Negative pacers
1
Amplitude Width
-2 to -700 mV 0.1, 0.2, 0.5 and 1.0 ms
-2 to -500 mV 1.5 ms
-2 to -400 mV 2 ms
1
Philips does not claim, verify, or validate
support for all available pacemakers.
EMC Performance
Limits, radiated
immunity
Meets Essential Performance, but may observe
some waveform disturbance over 181 to 202 MHz
@ >1.8 V/m and 203 to 213 MHz @ > 1.0 V/m
Parameter
Specification
Summary of Contents for IntelliVue TRx
Page 6: ...Draft 1 Aug 08 About this Book vi ...
Page 40: ...Draft 1 Aug 08 Accessories 2 18 Product Safety ...
Page 54: ...Draft 1 Aug 08 Audible Tones 3 14 Transceiver Controls ...
Page 74: ...Draft 1 Aug 08 Battery Information 4 20 Basic Operation ...
Page 94: ...Draft 1 Aug 08 Technical Alarms INOPs 5 20 Alarms ...
Page 152: ...Draft 1 Aug 08 How the QT Analysis Algorithm Works 7 24 ST AR Arrhythmia Monitoring ...
Page 176: ...Draft 1 Aug 08 Optimizing SpO2 Measurement Performance 8 24 SpO2 Monitoring ...
Page 236: ...Draft 1 Aug 08 Troubleshooting 11 16 Maintenance Cleaning Troubleshooting ...
Page 268: ...Draft 1 Aug 08 SpO2 Accessories A 10 Accessories ...
Page 270: ...Draft 1 Aug 08 B 2 Sales and Support Offices ...
Page 274: ...Draft 1 Aug 08 Index 4 ...