Draft - 1 Aug 08
Testing Transceiver Functionality
4-12
Basic Operation
Status
Check
You can check the status of the transceiver indicators at any time.
Initiating a
Status Check
Step
Action
1
Press the
Check
button.
The following indicators should illuminate for as long as the Check
button is depressed.
•
Battery gauge
•
Leads Off indicators
•
EASI (if in use)
2
If one or more of the expected indicators do not light up, check the
following:
•
Power and position of batteries (see “Checking the Battery
Power Level” on page 4-17).
•
Leadset insertion. Make sure the leadset is correctly inserted
in the transceiver and the colored line at the base of the cable
is not visible (see “Connecting the ECG Cable” on page 6-
21).
•
Lead positions and connections (see “Verifying Electrode
Connections” on page 6-25).
If there is still a problem, contact your service provider for
assistance.
Summary of Contents for IntelliVue TRx
Page 6: ...Draft 1 Aug 08 About this Book vi ...
Page 40: ...Draft 1 Aug 08 Accessories 2 18 Product Safety ...
Page 54: ...Draft 1 Aug 08 Audible Tones 3 14 Transceiver Controls ...
Page 74: ...Draft 1 Aug 08 Battery Information 4 20 Basic Operation ...
Page 94: ...Draft 1 Aug 08 Technical Alarms INOPs 5 20 Alarms ...
Page 152: ...Draft 1 Aug 08 How the QT Analysis Algorithm Works 7 24 ST AR Arrhythmia Monitoring ...
Page 176: ...Draft 1 Aug 08 Optimizing SpO2 Measurement Performance 8 24 SpO2 Monitoring ...
Page 236: ...Draft 1 Aug 08 Troubleshooting 11 16 Maintenance Cleaning Troubleshooting ...
Page 268: ...Draft 1 Aug 08 SpO2 Accessories A 10 Accessories ...
Page 270: ...Draft 1 Aug 08 B 2 Sales and Support Offices ...
Page 274: ...Draft 1 Aug 08 Index 4 ...